Nova-tech pH 310 Instrukcja Użytkownika Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 17
3534
P4.3 Calibration buffer selection sets
Available in pH 310 meter only
The pH 310 meter lets you select between three standard calibration buffer sets,
depending on your requirements. The available sets are USA, NIST, and DIN standard
calibration buffers.
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
su/jointfilesconvert/1428084/bgroups until you view parameter P4.0.
See figure
3. Press the ENTER key three times to
select parameter 4.3.
See figure
4. Press the and keys to select the
buffer set you require:
• USAbuffers
(pH 1.68, 4.01, 7.00, 10.01, 12.45)
• NIST buffers
(pH 1.68, 4.01, 6.86, 9.18, 12.45)
• DIN buffers
(pH 1.09, 3.06, 4.65, 6.79, 9.23, 12.75)
5. Press the ENTER key to confirm selection
and to return to the su/jointfilesconvert/1428084/bgroup menu.
Press the CAL/MEAS key to return to
measurement mode.
P4.4 Selecting °C or °F
Available in pH 310 meter only
The pH 310 meter lets you select between °C and °F units for temperature readings.
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
su/jointfilesconvert/1428084/bgroups until you view parameter P4.0.
See figure
3. Press the ENTER key four times to
select parameter 4.4.
See figure
4. Press the and keys to toggle
between °C and °F.
5. Press the ENTER key to confirm selec-
tion and to return to the su/jointfilesconvert/1428084/bgroup menu.
Press the CAL/MEAS key to return to
measurement mode.
SETUP
SETUP
E
F
SETUP
SETUP
F
G
G
H
H
E
F
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 24

Komentarze do niniejszej Instrukcji

Brak uwag